Make sure the tape has not been dented or kinked. This may
affect the accuracy of the tape.
Pi Tape gages are invaluable tools to the aero-space field as
well as the plastic pipe industry. The variety of uses are infinite.
EXAMPLE
Each line on the Gage Member of the tape represents .5mm of diameter,
while each line on the Vernier represents .01 mm. The zero on
the Vernier is past the 661.5 line on the Gage Member of the tape.
Now find the line on the Vernier which corresponds exactly with
the line above. The example shows it to be the 23 line which equals
.23mm. This added to the 661.5 reading gives an outside diameter
of 661.73mm.
When using standard O.D. tape on an I.D. surface, add double the
tape thickness to the reading to arrive at the l.D. of the part.
It is suggested that direct I.D. tapes be used for Inside Diameter
readings.
As a suggestion for checking very large diameters - pieces of
masking tape may be used to hold the tape in proper parallel position.
When reading the O.D. tape, apply a snug pull of 5 pounds tension,
first making sure the tape and part have been properly cleaned.
Pi Tapes Gages are guaranteed +/- .03mm accuracy on standard tapes
up to 3600mm.
Care
When not in use, wipe clean and apply a light rust preventive
oil. Store in tape container.
No periodic adjustments are needed.
Make sure the tape has not been stepped on or kinked, which may
destroy the accuracy.
Back to Top of Page